ISO 5660 heat capacity of cone calorimeter

2023-02-22

iso 5660 (1).png

  Introduction:  

Cone calorimeters use the oxygen consumption concept to calculate the rate at which materials emit heat. According to the oxygen consumption principle, every unit of oxygen consumed when a substance burns produces essentially the same amount of heat.


Cone Calorimeter can measure the sample's mass loss rate, heat release rate, smoke generation rate, ignition time, oxygen consumption status, and carbon monoxide and carbon dioxide generation when the sample is exposed to a cone heater's heat source.



  Application:  

Cone calorimeters use the oxygen consumption concept to calculate the rate at which materials emit heat.

According to the oxygen consumption principle, every unit of oxygen consumed when a substance burns produces essentially the same amount of heat.



Cone Calorimeter can measure the sample's mass loss rate, heat release rate, smoke generation rate, ignition time, oxygen consumption status, and carbon monoxide and carbon dioxide generation when the sample is exposed to a cone heater's heat source.



  Standards:  

• ISO 5660

Reaction to fire tests – Heat release, smoke production and mass loss rate 

• ASTM E 1354

Standard Test Method for Heat and Visible Smoke Release Rates for Materials and Products Using an Oxygen Consumption Calorimeter 

• BS 476 Pt.15

Fire tests on building materials and structures – Method for measuring the rate of heat release of products 



  Features:  

01/Test area equips with tempered glass for observation and protection during testing.

02/Brass calibration burner by clip-on installation method is used to calibrate the power of radiation cone.

03/The bottom of sample holder is equipped with imported weighing sensor for timely monitoring the weight loss.

04/Calibration burner adopts MFC to control the flow rate. MFC range 10L/min; accuracy ±1%F.S.; linearity ±0.5%F.S.; response time 2 sec

05/Rated power of radiation cone: 5000W; heat output: 0~100 kW/m2

06/Spark igniter is convenient to measure burning time.

07/Imported heat flow meter, range of 0~100 kW/m2, accuracy ±3%, repeatability ±0.5%; with calibration report; Portable cooling water source is provided, no need external water source.

08/Imported paramagnetic O2 analyzer and infrared CO2 analyzer; Oxygen range: 0~25%, accuracy <0.02%, repeatability <0.01%; CO2 range: 0~10%, accuracy 0.1 ppm; CO range: 0~1%, accuracy 0.1 ppm, repeatability 0.1 ppm; Response time < 8s

09/Specimen gas pretreatment system consists of suction pump, various levels of filtering devices, condensers, flow limiting and splitting devices for providing stable specimen gas to analyzer.

10/Equipped with laser transmitter and silicon optical amplifier for measuring gas density in the tube, and have optical path correction device.

11/Several combustion parameters: heat release rate (HRR), total release heat (THR), effective heat of combustion (EHC), ignition time (TTI), Mass loss rate (MLR), smoke generation rate (SPR), total smoke generation, CO2/CO generation.